IRS-1

Product Image
Product Overview

It is applicable in the industry such as electrical power, coal, commerce inspection, environment protection, metallurgy, paper making, petrochemical, earth probe, and the research institution or quality inspecting organization to measure the total sulfur content in the materials of coal, cinder, coke, catalyst, minerals, rock, and petrochemicals.

  • Measurement range: 0.01% ~ 40% (can be expanded according to user requirements)
  • Test method: infrared spectrum absorption method
  • Sulfur measurement resolution: 0.001%
  • Working furnace temperature: 1300 ℃
  • Temperature control accuracy: ± 1 ℃
  • Sample Weight: 300 ± 10mg (coal, coke), 100 ± 10mg (petroleum products)
  • Sample sending method: automatic sample sending, discarding sample
  • Number of samples: 12 / batch, test samples can be added / replaced
  • Single sample analysis time: about 2min
  • Accuracy: In accordance with GB / T25214-2010 "Determination of total sulfur in coal by infrared spectroscopy"
  • Power: ≤4kW
  • Working power: AC220V ± 22V / 50Hz
  • Dimensions (mm): 670 × 580 × 720
  • Overall weight (kg): 87
Note: Due to continuous improvements specifications given above can be changed without prior notice